1. Introduction
Congratulations on your selection of the STATIM Cassette Autoclave
We are confident that you have purchased the finest equipment of its type. The STATIM
unit that features a number of sterilizing cycles designed to meet your needs and suitability for steam sterilization.
The details of installing, operating and servicing your STATIM are all contained within this operator's manual.
To ensure years of safe, trouble-free service please read these instructions before operating this unit and keep them
for future reference. Operational, maintenance and replacement instructions should be followed for the product
to perform as designed. Contents of this manual are subject to change without notice to reflect changes and
improvements to the STATIM product.
The STATIM is suitable for the sterilization of dental and medical instruments designed to withstand steam sterilization.
The STATIM is not designed to sterilize liquids, cloth loads, biomedical waste or materials not compatible with steam
sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.
For more information about instrument suitability for steam sterilization, consult the manufacturers' reprocessing
instructions.
1.1 Disclaimers
Use only steam-process distilled water in your STATIM. Deionized, demineralized, or specially filtered water should
not be used. Never use tap water.
Do not permit any person other than certified personnel to supply parts for, service or maintain your STATIM.
SciCan shall not be liable for incidental, special or consequential damages caused by any maintenance or services
performed on the STATIM by a third party, or for the use of equipment or parts manufactured by a third party,
including lost profits, any commercial loss, economic loss, or loss arising from personal injury. Never remove the
cover of the unit and never insert objects through holes or openings in the cabinetry. Doing so may damage the unit
and/or pose a hazard to the operator.
All elements of this book are common to both STATIM 2000 and STATIM 5000 except when noted.
IMPORTANT:
Drying Performance
The STATIM 2000 and 5000 have been designed to provide a complete sterilization solution for your unwrapped
and wrapped instrument needs: rapid sterilization balanced with rapid drying, by using the SciCan Dri-Tec Drying
Technology.
To dry instruments, the STATIM 2000 uses a combination of forced filtered air and convection heat. The convection
heat is derived from utilizing the remaining heat in the system after the sterilization phase has been completed.
The heat is then captured and released into the cassette to rapidly dry a properly loaded cassette.
The STATIM 5000 also uses forced filtered air and heat generated by the sterilization phase to dry the instruments.
The heat generated by the sterilization phase is absorbed by the drying plates and then transferred directly to the
load. This then results in an accelerated, rapid drying of a properly loaded STATIM cassette.
Please refer to this operator's manual for instructions on the proper arrangement of instruments in the cassette and
the use of STAT-DRI™ plates (STATIM 5000). By carefully following these directions on how to properly load the
cassette chamber, rapid drying of the load will be achieved.
SciCan Ltd.
Introduction
Follow local guidelines governing verification of the sterilization procedure.
®
.
2
STATIM 2000/5000 Cassette Autoclave Operator Manual
®
is a compact, counter-top