3.3 Connecting the Waste Bottle

The waste bottle
converted to steam and then drained from the cassette. To connect the
waste bottle to the STATIM, follow these steps (see Figure 4).
1.
Insert the exhaust
and connect tightly.
2.
Cut the tube to length and slide the waste bottle fitting
3.
Place the free end of the tube into the hole in the lid of the waste bottle
and hand-tighten the fitting. Do not coil the exhaust tube.
4.
Unscrew the lid and copper condenser coil
waste bottle. The lid and coil should come out together.
5.
Fill the waste bottle with water to the MIN line and replace the lid and
copper condenser assembly. Empty the waste bottle often to avoid
unpleasant odors and discoloration of the contents. (A low-level
disinfectant, prepared according to the manufacturer's instructions,
may be added to the waste bottle to remedy this situation). As a
minimum, empty the waste bottle each time you refill the reservoir.
6.
Place the waste bottle near the unit. Store the bottle below the unit.
The tube can be routed through a hole, (8 mm/0.3˝ in diameter) in the
counter-top and secured with the provided nylon clamps.

3.4 Filling the STATIM Reservoir

When filling the reservoir, ensure you only use steam processed
distilled water containing less than 5 ppm total dissolved solids (having
conductivity of less than 10 µS/cm). The impurities and additives in other
water sources will cause an error reading on the LCD. If you have a water
conductivity meter (available from SciCan, order number 01-103139S)
check each new water container before filling the reservoir. If your unit is
equipped with a Rev 7 PCB or higher (software SxxxR6xx), you can read
the water conductivity by refering to Section 5.9 Reading Water Quality.
To fill the reservoir, follow these steps (see Figure 5):
1.
Remove the reservoir cap (2).
2.
Pour steam-process distilled water into the reservoir until almost full
(a maximum of 4L /1 U.S. gal). Use a funnel to avoid spillage.
3.
Replace and secure the cap.

3.5 Priming the STATIM Pump

To prime the STATIM pump, follow these steps (see Figure 6):
1.
Move the unit to the edge of the work surface. The front leveler feet
should be approximately 12 mm/0.5˝ from the edge.
2.
Lift the front left corner of the unit upward and remove the drain tube
(3)
from the clip located on the underside of the unit.
3.
Pull the drain tube outward so the free end can be positioned over a
water container.
4.
Fill the reservoir with steam-process distilled water.
5.
Remove the plug
to drain from the tube into a container for 30 seconds. When the water
flows in a steady stream, replace the stopper.
6.
Lift the front left corner of the unit upward and reinsert the tube into
the clip on the underside of the unit. Push the excess length of tubing
back into the space provided.
Make sure the plug in the drain tube is secured.
After installation, and before sterilizing any instruments, run two Wrapped cycles.
For further instructions, see Section 3.10 Preparing Unit for Use.
SciCan Ltd.
Installation
is used to collect the wastewater after it has been
(2)
tube into the fitting
(3)
from the end of the drain tube
(4)
on the back of the unit
(4)
into place.
(5)
assembly from the
(6)
and allow water
(3)
6
3
6
2
STATIM 2000
3
4
2
2
4
3
STATIM 2000/5000 Cassette Autoclave Operator Manual
STATIM 5000
4
5
3
6
Figure 4
Figure 5
Figure 6