Electromagnetic Immunity

The Chorus Sound Processor is intended for use in the electromagnetic
environment specified below. The customer or the user of the sound
processor should assure that it is used in such an environment.
Immunity/Compliance
Standard
Electrostatic discharge
(ESD)
IEC 61000-4-2
Power Frequency
Magnetic Field
IEC 61000-4-8
18
Compliance level
a
Contact:
± 8 kV
Air:
± 15 kV
30 A/m
@ 50/60 Hz
Electromagnetic environment - guidance
If used when the operator may come
in contact with synthetic materials
that may harbor static charge (e.g.,
nylon carpeting or clothing), relative
humidity should be at least 10%. As
with the operation of other electronic
devices, precaution should be taken
to not generate ESD.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.