Attaching and removing the mini-dock
WARNING
To avoid the risk of electrical shock to the patient, do not touch the ungrounded
connectors on the back of the mini-dock while simultaneously touching a patient.
CAUTION
• To avoid damaging the mini-dock, do not perform high current grounding impedance
testing involving this part. Accessible metal of the mini-dock is not protectively
earthed.
• To avoid damaging equipment, remove mini-dock from system prior to transporting.
1.
Ensure that the ultrasound system is turned off.
Connecting the system and mini-dock while the power is on may cause the mini-dock to function
improperly.
2.
Slide the mini-dock feet under the ultrasound system, and firmly press the mini-dock into the system
connector. The mini-dock is correctly attached when no gap is between the system and mini-dock.
NOTE
If at any time the ultrasound system loses connectivity, turn off the system
and reattach the mini-dock and system. Always verify that the mini-dock and
system are correctly attached after changing a transducer. When you change a
transducer, the mini-dock may loosen.