Withings BPM Connect Manual De Instrucciones página 24

Ocultar thumbs Ver también para BPM Connect:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Declaration – electromagnetic emissions and immunity
for equipment and systems that are not life-supporting and are specified for use only in a shielded location
Declaration – electromagnetic immunity
The Wi-Fi Smart Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the Wi-Fi Smart Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
Contact: ±8 kV
(ESD)IEC 61000-4-2
Air: ±2 kV, ±4 kV, ±8 kV,
±15 kV
Electrical fast transient/
2 kV for power supply lines
burst IEC 61000-4-4
1 kV for input/output lines
Surge IEC 61000-4-5
1 kV differential mode
2 kV common mode
Compliance level
Electromagnetic environment – guidance
Contact: ±8 kV
Floors should be wood, concrete or ceramic tile. If floors are
Air: ±2 kV, ±4 kV, ±8 kV,
covered with synthetic material, the relative humidity should
±15 kV
be at least 30 %.
N/A
Mains power quality should be that of a typical commercial
N/A
or hospital environment.
N/A
A Mains power quality should be that of a typical commer-
N/A
cial or hospital environment.

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para Withings BPM Connect

Tabla de contenido