Idiomas disponibles

5.7. Reading Recall - Displaying the last measurement

The measured results are stored in the instrument. This unit
stores 30 measurements. You can call up the stored values at
any time by pressing the MEMORY button. The first memory dis-
played will be the most recent measurement taken. Keep push-
ing the button to scroll back through memories.
Further information
Measurements should not occur soon after each other, to ensure accurate readings. Wait at least
1 minute in a relaxed position, sitting or lying, before you repeat a measurement.

6. Error messages/Troubleshooting

If an error occurs during a measurement, the measure-
ment is discontinued and a corresponding error code is
displayed.
(example: error no. 1)
Error No.
ER 1
ER 2
ER 5
10
Possible cause(s)
The systolic pressure was determined but afterwards
the cuff pressure fell below 20 mmHg. The tube may
have become unplugged.
Further possible cause: No pulse was detected.
The pressure dropped below 20 mmHg without a mea-
surement taking place. The Release valve was pressed
or the arm moved during the measurement.
Unit detected an unacceptable difference between sys-
tolic and diastolic pressures. Take another reading.
Contact your doctor if this continues.
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PULSE
/min
MR
SYSTOLIC
mmHg
DIASTOLIC
mmHg
PULSE
/min