Beurer IH 60 Instrucciones De Uso página 21

Inhalador
Ocultar thumbs Ver también para IH 60:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
Preparation
• Immediately after each treatment, all parts of the atom-
iser [9] and the accessories used must be cleaned of
residual medication and contamination.
• Remove the mouth piece [12], the mask [13, 14] or the
nosepiece [15] from the atomiser.
• Disassemble the nosepiece if you have used it with the
comfort attachment.
• Dismantle the atomiser by twisting the
top anticlockwise against the medica-
tion container [10].
• Remove the atomiser insert [11] from
the medication container [10].
• Reassembly is carried out in reverse
order.
Cleaning
Important
The device must be switched off, disconnected from
the mains and allowed to cool down each time before
cleaning.
The atomiser and the accessories used such as the
mouthpiece, mask, etc. must be washed with hot but not
boiling water after each use. Dry the parts carefully using
a soft cloth. Put the parts together again when they are
completely dry and place them in a dry, sealed container
or disinfect them.
When cleaning, ensure that any residue is removed. Never
use any substances for cleaning that could potentially be
toxic if they came into contact with the skin or mucous
membranes, or if they were swallowed or inhaled.
Use a soft, dry cloth and non-abrasive cleaning products
to clean the device.
Do not use any abrasive cleaning products and never
submerge the device in water.
Important
• Ensure that no water gets inside the device.
• Do not clean the device or accessories in the dishwash-
er.
• Do not touch the device with wet hands when it is
plugged in and do not allow water to spray on the de-
vice. Only operate the device if it is completely dry.
• If liquid penetrates the device this could cause damage
to the electrics or other nebuliser parts and lead to a
malfunction.
Condensation, hose care
Condensation may form in the hose depending on the
ambient conditions. It is essential to remove the moisture
to prevent bacterial growth and ensure proper treatment.
To do so, proceed as follows:
10
• Remove the compressed air hose [8] from the atomiser
[9].
• The hose must remain connected to the nebuliser [6].
11
• Operate the nebuliser until the moisture is removed
by the air passing through.
• In the event of heavy contamination, replace the hose.
Disinfection
Please carefully observe the points below when disin-
fecting your atomiser and accessories. We recommend
disinfecting the individual parts on a daily basis after the
last usage as a minimum measure.
(All you need for this is a little white vinegar and distilled
water.)
• First, clean the atomiser and accessories as described
in the "Cleaning" section.
• Place the disassembled nebuliser, mouthpiece, and
nosepiece in boiling water for five minutes.
• For the remaining accessories, use a vinegar solution
consisting of ¼ vinegar and ¾ distilled water. Make
sure that the volume is sufficient to fully submerse the
parts, such as the atomiser, mask and mouthpiece, in
the solution.
• Leave the parts in the vinegar solution for 30 minutes.
• Rinse the parts with water and dry them carefully with
a soft cloth.
Important
Do not boil or autoclave the compressed air hose or the
masks.
• Put the parts together again when they are completely
dry and place them in a dry, sealed container.
21

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido