LOCK
fill set
UNLK
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch
to the next setting:
"UNLK" the keypad (-) or "LOCK" the keypad (+). In the "LOCK" mode the feeding program of the pump can no longer be
changed.
Use the "+" or "-" key to change the setting from "LOCK" to "UNLK" or vice versa.
+
BEEP WHEN DONE
fill set
MUTE WHEN DONE
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch
to the next setting:
"BEEP WHEN DONE" (+): the pump will give an alarm when the dose is administered or "MUTE WHEN DONE" (-): the pump
will not give an alarm after administering the dose. Use the "+" or "-" key to change the setting from "BEEP WHEN DONE" to
"MUTE WHEN DONE" or vice versa.
+
LITE ON
fill set
LITE OFF
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch
to the next setting:
"LITE ON" (+) the light of the LCD remains on when the pump is connected to the mains. "LITE OFF" (-) the light of the LCD
switches off after a few seconds even if connected to the mains. Use the "+" or "-" key to change the setting from "LITE ON" to
"LITE OFF" or vice versa.
+
OUTP PMDS
fill set
OUTP NRSE
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the "FILL SET" key to switch
to the next setting:
"OUTP PDMS" (+) to connect with a Patient Data Monitoring System. "OUTP NRSE" to connect with a NURSE CALL system. Use
the "+" or "-" key to change to setting from "OUTP PDMS", to "OUTP NRSE" or "OUTP OFF" and vice versa.
Press the
key and keep it pressed for 2 seconds to exit the SET UP mode and turn off the pump. The settings are automa-
tically saved.
402981282_r1.indd 98
402981282_r1.indd 98
TO LOCK THE KEYPAD
TO MUTE WHEN DOSE DONE
TO SWITCH LIGHT ON
TO CONNECT WITH NURSE CALL OR PDMS
2 sec
-98-
EN
06/01/23 4:50 AM
06/01/23 4:50 AM