Instructions for use elements e-motion Dental Motor
6 Operation | 6.1 Switching the elements e-motion on/off
6 Operation
Incorrectly set input values.
Property damage from incorrect input values.
▶ Check the following input values on the display before use:
▪ Speeds
▪ Transmission ratios
▪ Torque
▪ Direction of motor rotation
▪ Settings
▪ Own settings
Note
Please note the transmission/reduction ratio of the attached handpieces as
these have an impact on the displayed speed and the display torque of the
clamped tool.
Note
If the foot control is pressed continuously without interruption, e.g. when
something is lying on the foot control, both the LUX light and the motor are
automatically switched off after approx. 10 minutes.
6.1 Switching the elements e-motion on/off
The unit is ready for operation as soon as it is connected.
▶ To switch the product on, connect the device to the power circuit.
ð
The display of the elements e-motion shows the input values:
Display elements e-motion
▶ To switch the device off, unplug the unit from the mains.
Note
The power consumption on idling is so low that the device does not need to
be disconnected.
6.2 Stand-by mode
The device switches to a screen saver mode if it is not used for a period of 10
minutes.
The display is switched off to save power if the device is not used for a period of
30 minutes.
▶ To reactivate the display, touch the display or press the foot control.
NOTICE
30 / 58