ERROR INFORMATION
Err0: The calibration data should not less than 4 digits.
Or the data will be invalid.
Err1: The calibration data should not exceed the data 65536.
Err2: There is incorrect password to modify the calibration data.
Err3: The weight value at the calibration point is too small.
It should be no less than 3 valid digits.
Err4: Incorrect calibration password. Please input the correct password.
Err5: The internal code at the calibration point is too small.
The valid internal code is from 3000 to 65536.
FAULT ANALYSIS BY CHECKING THE INTERNAL CODE
Press [Zero] to power on the scale, it will first show "Ad". Then un-press the key,
it will show the internal code. If there is no load on the scale, the internal code
should be between 6000 and 8000. If the internal code is not in the range from
6000 to 8000, please send the scale back to your supplier for repair.
*
The products are subject to change without notice.
EN
29
www.steinbergsystems.com