5.2 Frequency measurement
To perform frequency measurement:
set the rotary switch at V
press MODE
connect black test lead to COM terminal, and red test lead to
VΩHz%
contact the tips of test probes to the points of measurement,
read the measurement result on the display.
5.3 Measurement % of duty cycle (pulse filling indicator)
To perform the measurement:
set the rotary switch at V
press MODE
connect black test lead to COM terminal, and red test lead to
VΩHz%
contact the tips of test probes to the points of measurement,
read the measurement result on the display.
5.4 Measurement of resistance
WARNING
Do not perform measurements on the circuit under the
voltage. Before the measurement disconnect the pow-
er and discharge capacitors.
To perform measurement of resistance:
set the rotary switch at Ω
connect black test lead to COM terminal, and red test lead to
VΩHz%
contact the tips of test probes to the points of measurement; the
best solution is to disconnect one side of the tested element, to
prevent the remaining part of the circuit interfere with the read-out
of the resistance value,
read the measurement result on the display.