Dismounting, Return And Disposal; Return - WIKA CPP700-H Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

8. Dismounting, return and disposal

10. Rinse the fluid reservoir using water or gentle cleaning agents.
To do so, use the syringe specially designed for cleaning.
EN
The use of alcohol, spirit, acetone or benzol can cause the container
material to crack or lead to damage of the container.
11. Screw in the plug screws
If water is used as test or calibration medium, the test pump must be
emptied after EACH use.

8.2 Return

Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances
(acids, bases, solutions, etc.) and must therefore be cleaned before being returned.
WARNING!
Physical injuries and damage to property and the environment
through residual media
Residual media at the hand test pump can result in a risk to persons, the
environment and equipment.
When returning the instrument, use the original packaging or a suitable transport
packaging.
To avoid damage:
1. Place the instrument, along with the shock-absorbent material, in the packaging. Place
shock-absorbent material evenly on all sides of the transport packaging.
2. If possible, place a bag, containing a desiccant, inside the packaging.
3. Label the shipment as transport of a highly sensitive measuring instrument.
Information on returns can be found under the heading "Service" on our local
website.
22
again.
3
With hazardous substances, include the material safety data sheet for the
corresponding medium.
Clean the instrument, see chapter 7.2 "Cleaning".
WIKA operating instructions, models CPP700-H and CPP1000-H

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cpp1000-h

Tabla de contenido