Maximum / Minimum:
Displays maximum / minimum pressure values;
this is initiated upon powering the unit or since
the values were cleared.
1. When in Menu mode press
to scroll to 'Max' (maximum display); this
is indicated in small font to the left of the
display's unit of measure.
2. To clear both Min and Max values, press
➠
and hold
zero
3. Release of this key will leave you
in Menu mode. Use
continue scrolling through the menu
options, OR hold,
to Measurement mode.
Note: Clearing Minimum / Maximum values
will result in a blank display except for the
associated Min. / Max. icon and battery
indicator.
Note: The following actions will also clear
min/max values. Power off , zero gauge
function, tare function, field calibration
function, units programming, update rate, or
reset.
➟
▴
▾
,
➟
▴
▾
,
to
➠
to return
menu
Timer:
Controls how long the gauge will remain
powered ON once the Power key
pressed.
1. When in Menu mode press
to scroll until "TIMER" is displayed.
➟
|
|
2. Press
3. Then, press
through values. "NONE" designates
that the gauge will remain ON until
the power key
time. Any other value (1 min [default]
/ 5 min / 20 min), other than "NONE",
will designate the duration of time this
function will be in eff ect.
|
|
➟
4. Press
timer icon will be shown on the display
and the unit will display "DONE" then
return to Measurement mode.
Feature: Timer icon will fl ash 10 seconds
prior to gauge shut down. While the
gauge is powered on, any key press will
reset the timer.
9
|
|
▴
▾
,
.
➟
▴
▾
,
to scroll
|
|
is pressed a second
to select desired value;
is
➟