_______________________________________________________________________________
5. After recharging, LCD displays "Power: 100%".
6. Remove the AC Adaptor.
1. Do not use non-rechargeable batteries to perform
charging.
2. Users are suggested to do recharging only when
LCD displays the Power is
3. Do not stop recharging when it is not finished so that
the battery life can be maintained.
Fuse Replacement
When the voltage can not be measured (Vnow = 0V) after properly connecting the Analyzer
and the solar panel, please check the fuse.
If the fuse is damaged (burned), please replace a new fuse by following the procedures:
1. Turn off the Analyzer and remove all the connecting wires and power sources.
2. Unscrew the battery cover. Remove the battery cover. Remove the two batteries on the
furthest east and west.
3. Unscrew the (4pcs.) screws of the bottom cover. Remove the bottom cover. Remove the
power connector connecting the bottom cover and the circuit board (J2).
4. Remove the damaged (burned) fuse.
5. Put in a new fuse of the same specifications (5A / 250V).
6. Connect the power connector. Replace and screw the bottom cover.
7. Replace the two batteries on the furthest east and west. Replace and screw the battery
cover.