a. DPD-1 test. Take a water sample from the small plastic tap
placed at the bottom part of the probe holder.
b. Chlorine Menu -> Select "calibration" and press OK. Wait until
the reading is stable:
c. Adjust the real value obtained by means of DPD-1 test.
d. Return to the main screen. The current CL value will match
with the value obtained from DPD-1 test.
3.2.5.3.5 Probe maintenance
Please, carefully read the following maintenance instructions for your PRO/2 Kit.
The maintenance periods are illustrative, they are depending on factors such
as the quality of the water and the installation maintenance, especially with
the filtration system.