Apéndice C -
The Intraoral Scanner is intended for use in the electromagnetic environment specified below. The user of the Intraoral
Scanner should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Tabla 1 
The Intraoral Scanner is intended for use in the electromagnetic environment specified below. The user of the Intraoral
Scanner should assure that it is used in such an environment.
Immunity Test
IEC 60601 test level
± 8 kV contact
Electrostatic
discharge (ESD)
± 2/4/8/15 kV air
IEC 61000-4-2
 
Electrical fast
± 2 kV
transient/burst
100 kHz repetition
IEC 61000-4-4
frequency
± 0.5/1 kV line to line
Surge
± 0.5/1/2 kV line to
IEC 61000-4-5
ground
Apéndice C - Directriz de compatibilidad electromagnética
Directriz de compatibilidad electromagnética
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
± 8 kV contact
± 2/4/8/15 kV air
 
± 2 kV
100 kHz repetition
frequency
± 0.5/1 kV line to line
± 0.5/1/2 kV line to
ground
Electromagnetic environment – guidance
The Intraoral Scanner uses radio frequency (RF) energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment.
The Intraoral Scanner is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low- voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.