2. Description of the Product
Pentamix
2, manufactured by 3M ESPE, is an automatic mixing unit for
processing of Penta impression materials. Penta impression materials,
manufactured by 3M ESPE, were specifically designed for this device
and are supplied in specialized foil bags. Only this form of packaging
permits the use in the Pentamix device. Each Penta foil bag is sealed
with a PentaMatic sealing cap. The external geometry of the sealing
cap ensures proper positioning on the cartridge. After the Pentamix
device is started up, a special internal device opens the foil bag once
sufficient pressure is established by the plunger. The connection pieces
located on the PentaMatic sealing caps serve for proper attachment
of the mixing tip.
The cartridges, labeled like the product, and the PentaMatic sealing
caps are the components of a mechanically sound unit for processing of
pastes in the Pentamix device that allows a change of products without
intermediate cleaning of the device and no loss of paste.
The Pentamix is supplied as a table-top unit, but can be converted into a
wall-mounted unit with the respective wall mounting brackets, which can
be ordered from 3M ESPE (Art. No. 077600).
3. Technical Data
Drive unit:
2 motors with built-in automatic thermostat
Operating mode: Short term operation - max.1.5 min. on, min.10 min. off
(prevents overheating of the electrical elements)
Cartridges:
Containers for Penta foil bags.
Dimensions:
60 x 130 mm, 26.8 x 130 mm
2.4 x 5.1 inches,
1.1 x 5.1 inches respectively
Housing:
All parts of the housing are made of impact-resistant
polycarbonate
Power supply:
120 V 230 V Please refer
Frequency:
60 Hz 50 Hz
Power input:
2.7A 1.5 A
Ambient
temperatures:
18°C/64°F.....40°C/104°F
Rel. humidity
of air:
20% ..... 80%
Classification:
Hazard class I
Dimensions:
370 mm x 240 mm x 307 mm
(L x W x H):
13.9 x 6.6 x 8 inches
Weight:
10.1 kg, 22,3 lbs
Continous sound
pressure level:
>64 dB(A)
Year of
manufacture:
See rating plate
4. Installation
! !
Prior to installing the unit, make sure
the power supply corresponds to the supply
quoted on the unit.
! !
In order to work with the unit, the power plug must be connected into
a suitably grounded socket. In case of using an extension cable make
sure that the ground line is not interrupted.
Pentamix 2 can be placed on any non-skidding flat surface. After
plugging in the powerline the unit is ready for use.
6
to rating plate
for voltage and
frequency.
The power cable (after having unscrewed the base of the unit), can be
wound around the cable hooks at the bottom of the unit.
Pentamix 2 can also be mounted to a suitable wall. A special wall fixture
is available and can be ordered from 3M ESPE (Art. No. 077600).
5. Functioning
5.1 Loading the Cartridge
! !
Please carry out the procedures in
the following sequence.
Prepare the foil bags of base paste and catalyst paste and the
cartridge of the product to be used.
Insert the foil bags from the front into the cartridge (Pict.
Make sure that the locating lugs are in the correct position (Pict.
Lock the lever for the mixing tip (Pict.
5.2 Inserting the Cartridge
Open the access door of the Pentamix 2 (Pict.
Take the cartridge in your left hand.
Turn the plunger adjustment knob clockwise with your right hand until
it stops and hold it in place.The plungers are now in the uppermost
position (Pict.
).
Place the cartridge into the Pentamix 2 (Pict.
Rotate the plunger adjustment knob anticlockwise until the plungers
markedly touch the foil bags (Pict.
! !
The Pentamix 2 cannot be
activated until the plungers
touch the foil bags.
Attach a mixing tip prior to beginning the mixing process (refer to next
section).
5.3 Attaching the Mixing Tip.
Always attach a mixing tip before use!
! !
Foil bags equipped with a PentaMatic
sealing cap must be exclusively used with
Penta mixing tips.
Open the access door of the Pentamix 2.
Release the locking lever.
Push the mixing tip over the drive shaft and then into the corresponding
holes in the cartridges. It may be necessary to rotate the mixing tip
slightly on the drive shaft for correct seating (Pict.
Press the locking lever down and lock into place.
Close the access door of the Pentamix 2.
5.4 Starting a New Cartridge
When starting a new pair of foil bags, first of all the pressure in the
cartridge has to build up evenly.
Both foil bags open automatically after some 15-25 seconds due
to the pressure exerted by the plunger. This event is easy to hear.
Catalyst and base paste bags may not open at exactly the same
time.