DIGITAL CLINICAL THERMOMETER
GERATHERM© SolarTherm® "Speed" GT-161/1
Geratherm Medical AG
Germany
Please read carefully before using.
PECIFICATIONS
Specifications:
Range:
89.6° ~111.0°F ; Below 89.6°F, display L°F Above 111.0°F,
displays H°F
Resolution:
0.01°F
Accuracy:
±0.2°F (93.2°F to 107.6°F), ±0.4°F (other ranges)
Display:
Liquid crystal display 3 1/2 digits
Energy Supply:
Integrated solar cell
Power Consumption:
0.15 milliwatts in measurement mode
Power Capacity:
Approximately 25 temperatures per 60-minute recharge
Dimensions:
5.79 in (147mm) x 0.84 in (21.3mm) x 0.59 in (15mm)
Weight: Approx.
0.64oz (18.2 grams)
Beeper:
Beeps for 8 seconds after peak temperature is reached
Memory:
Built-in memory enabling display of the previously measured
value
Shut-off:
Automatic shut-off in abt. 9 minutes
Operating Conditions:
Temperature: 50°F ~ 95°F; Humidity: 30%~75%
Storage Conditions:
Temperature: 14°F ~140°F; Humidity: 10%~83%
Guarantee of Quality:
EN 12470, EN ISO 9001, EN ISO 13485, Council Directive
93/42/EEC
Precautions:
1. Do not boil or otherwise artificially heat the probe. Instead, clean the unit by wiping it with
a dry cloth and disinfect the probe with ethyl or Isopropyl (rubbing) alcohol.
2. Do not drop the thermometer or expose it to heavy shock.
3. Do not bend or bite the probe.
4. Do not store the unit under direct sunlight, at a high temperature, in high humidity or dust.
5. Stop using the thermometer if it operates erratically or if the display malfunctions.
6. Keep out of the reach of unsupervised children.
7. Do not attempt to disassemble the unit.
8. If the thermometer has been stored at below-freezing temperatures, allow it to warm
naturally to room temperature before using.
Power and Recharging:
1. For the first time use, the unit must be exposed to a light source at least 3 hours.
2. To properly charge, the light source must be 500 Lux or above.
3. When the low power symbol "
"appears, the solar cell needs to be recharged at least 60
minutes.
4. Turn the power off before recharging.
5. The solar cell should face the light source during recharging.
Directions for Use:
1. Disinfect the probe with ethyl or isopropyl (rubbing) alcohol before using.
2. Depress the ON/OFF button to activate the display.
3. Release the ON/OFF button and the display will show L°F with a flashing degree scale
symbol.
4. Place the probe correctly in the appropriate site (oral or rectal).
5. Once the degree scale symbol stops flashing, the correct temperature is displayed. The
temperature reading will not change after the degree sign stops flashing.
6. The unit will automatically turn off in 9 minutes (approx.). However, to save energy, it is
best to turn the thermometer off immediately after the temperature has been noted.
Special Features:
Normal Temperature Alarm: An alarm will sound when the maximum temperature has
registered and is ready to read. The "Normal Temperature" alarm (four quick beeps followed
by a pause) will sound repeatedly for approximately 8 seconds.
Fever Alarm: If the temperature registers higher than 99.5°F, the fever alarm (a quick
beeping sound, which occurs continuously for about 8 seconds), will sound when the peak
temperature has registered.
Last Memory Recall: When the ON/OFF button is depressed to activate the thermometer,
display of 188.8 will change to show the last measured temperature and then will return to
"ready-to-measure" mode. The recalled temperature will then be erased.
NOTE: Always disinfect the thermometer before use.
NOTE: If the room temperature is greater than 89.6°F, initial reading will be displayed as a
number rather than L °F.
NOTE: You may put a sanitary plastic probe cover on the sensor tip to help prevent cross-
infection. Use of the probe cover can result in a 0.2°F difference from actual temperature.
Dispose of any used probe covers and store the unit in a safe place.
ORAL
Place the sensor tip well under the patient's tongue. Instruct the patient to keep their mouth
closed while the thermometer is measuring.
RECTAL
If rectal temperature is recommended by your doctor, you should use a probe cover.
Insert thermometer into the probe cover. Lubricate the probe cover with a water soluble
jelly for easier insertion. Do not use petroleum jelly. Insert tip of probe no more than
1/2 inch into the rectum. STOP if you meet any resistance.
AXILLARY
Wipe armpit with a dry towel. Place the probe high in the patient´s armpit and keep the
patient´s arm pressed firmly against their side.
NOTE: Children should be attended by an adult during the entire temperature measurement
process. After use, the digital thermometer should be stored in a safe place, out of reach of
young children.
If you compare temperatures between two thermometers, it is possible to have two different
readings even if both thermometers are functioning within their accuracy tolerances. This can
be due to differences in technique.
REMEMBER: Find your normal temperature and watch for changes.
Cleaning Instuctions:
To clean the thermometer, wash the tip with a solution of mild detergent and cool water.
Disinfect the thermometer by wiping the sensor and lower stem with a cloth dipped in a
household anti-septic solution such as rubbing alcohol. Please make sure that the
thermometer is not totally immersed in water or disinfection solution. The thermometer is
water resistant, but not waterproof.
Limited Warranty
This thermometer is guaranteed for 2 years from the date of purchase against
manufacturer's defect under normal use. If your unit does not function properly due to
defective parts or assembly we will repair it free of charge. All parts are covered by this
warranty. The warranty does not cover damage to your unit due to improper handling.
The manufacturer recommends verification of accuracy by an authorized laboratory every 2
years.
Manufacturer:
Geratherm Medical AG
Fahrenheitstraße 1
D-98716 Geschwenda, Germany