4 Operation
4.4
Troubleshooting
4.5
Switching off
the unit
30
Error messages
Possible causes and solutions
Err 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse could be
determined.
Check the connections between the cuff and the main unit.
Err 2
Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
Err 3
If the cuff takes too long to inflate, it is possible that the cuff is incorrectly
positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
Err 5
An average value could not be calculated due to unstable conditions while the
readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.
Err 6
There were too many errors during the measurement in MAM mode, making
it impossible to obtain a final result.
Read through the checklist for performing reliable measurements and then
repeat the measurement.
HI
The pressure in the cuff is too high (over 300 mmHg) or the pulse is too high
(over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
LO
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
If no buttons are pressed for approx. 1 minute, the device automatically switch-
es to standby mode. You can also switch to standby mode by pressing the
START button
. The blood pressure meter functions are now switched off.
The time and the selected user memory are continuously shown on the display.
The alarm function remains active.
If you do not want to use the time display and the alarm function, you can also
completely switch off the device by sliding the ON/OFF switch
of the device into the " OFF " position.
on the back