7. Technical data

Technical data
Ambient conditions
Duringoperation:� � � � � � � � � � � � � � � � � � � � � � � � � � � �  T emperature10°to40°C,Humidity0to90%

Duringstorage:� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �  T emperature-20°to70°C,Humidity0to90%

Performance data
Voltage � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 110V/127V/230V;AC,50to60Hz
Fuses�����������������������������������������������������������IEC127-T800L250V,2EA
Powerconsumption:� � � � � � � � � � � � � � � � � � � � � � � Ultrasonicnebulizer50VA;Hoseheating30VA
Resistanceofhoseheating � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �3Ohm(±5%)
Ultrasonicfrequency � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 1,68MHz(±5%)
Noisepressurelevel � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � about35dB(A)
Humidityprotectiontype� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �IP32(drippingwater)
Method of operation: Continuous operation
Nebulizeroutput� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � max�3ml/min
Nebulizeroutput� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �withsterilewatersystemupto2,5ml/min
Dropletsize(MMAD) � � � � � � � � � � � � � � � � � � � � � � � � � � �average4µm,about86%lowerthan5µm
Airflow� � � � � � � � � � � � � � � � � � � � � � � � � � � � � �upto20L/min(whenusingAQUA+ForSecuRedBig)
Classification
asperECDirective93/42/EEC,AnnexIXClassIIa
CE-Labelling
Directive93/42/EECconcerningMedicalDevices
General information on electromagnetic compatibility (EMC) according to
the international EMC standard IEC 60601-1-2: 2001
Medicalelectricalequipmentneedsspecialprecautionsregardingelectromagneticcompa-
tibility(EMC)andneedstobeinstalledandputintoserviceaccordingtotheEMCinformation
providedinthetechnicaldocumentationavail-ablefromDeVilbissHealthcareServiceupon
request�
Portable and mobile RF communications equipment can affect medical electrical equip-
ment.
PinsofconnectorsidentifiedwiththeESDwarningsymbolshallnotbetouchedandnotbe
connectedunlessESDprecautionaryproceduresareused�Suchprecautionaryprocedures
mayincludeantistaticclothingandshoes,thetouchofagroundstudbeforeandduring
connectingthepinsortheuseofelectricallyisolatingandantistaticgloves�Allstaffinvolved
intheaboveshallreceiveinstructionintheseprocedures�
Dimensions:
Heightoverall� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 19,0cm
Width� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 20,5cm
Depht� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 31,5cm
Weight � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � about3,5kg
relativehumiditywithoutcondensation