Capacitance; Diode Threshold Voltage Measurement; Frequencies - Metrix MX 53C Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Chapter II
5.6.

Capacitance

"
Note
Discharge all capacitors before taking measurements.
Ranges
50 nF**
1% R + 2 D
500 nF
1% R + 2 D
5 µF
1% R + 2 D
50 µF
1% R + 2 D
500 µF
1% R + 2 D
5 000µF
2% R + 2 D
50 mF
2% R + 2 D
* Overload protection can be reset automatically.
**It is highly recommended to use very short and shielded test leads for measurements in
this range.
Number of points :
Range selection :
Maximum open circuit voltage :
5.7.

Diode threshold voltage measurement

Measurable voltages :
Measurement current :
Resolution :
Protection :
5.8.

Frequencies

Selector switch setting :
Measurement range :
Accuracy :
Protection :
Display :
RANGE
0.62 Hz to 5 kHz *
5 kHz to 50 kHz
50 kHz to 500 kHz
* Rectangular signals
Portable digital multimeter
Accuracy
Measurement
current
100 nA
1 µA
10 µA
100 µA
1 mA
1 mA
1 mA
Max measurement
time
0.5 s
0.5 s
0.5 s
0.5 s
1.5 s
3 s/mF
3 s/mF
5 000
automatic or manual
7 V
0 to 2 V
1 mA ± 20 %
1 mV
600 V
, can be reset automatically
RMS
V
+ SEL/ON
AC
0.62 Hz to 500 kHz
0.03 %
1100 V
PK
50 000 points
SENSITIVITY
5 V to 500 V
2 % of range
5 % of range
10 % of range
Protection*
Resolution
600 V
10 pF
RMS
600 V
100 pF
RMS
600 V
1 nF
RMS
600 V
10 nF
RMS
600 V
100 nF
RMS
600 V
1 µF
RMS
600 V
10 µF
RMS
750 V
100 V
250 V
////////////////
31

Publicidad

Tabla de contenido
loading

Tabla de contenido