en
5.3
Safe operation
Coagulation of patient tissue
or risk of burns for patients
and user through hot product!
WARNING
►
►
►
►
Risk of infection or injury
through aerosol formation or
particles ejected from the
WARNING
tool!
►
Risk of injury and/or malfunc-
tion!
►
WARNING
Risk of injury when attach-
ing/removing sawblades in the
ON position through inadver-
WARNING
tent activation of the product!
►
Risk of injury due to applica-
tion of the product outside the
field of view!
WARNING
►
12
Cool the tool during opera-
tion.
Keep product/tool out of
the reach of patients.
Let the product/tool cool
down.
Use a cloth to protect
against burns when chang-
ing the tool.
Use suitable protection
(such as waterproof pro-
tective clothing, face
mask, safety gases, suc-
tion).
Always carry out a func-
tion check prior to using
the product.
Only attach/remove saw
blades in the OFF position.
Only use the product under
vision.
Risk of injury and damage to
the tool/system!
The tool may get caught in
WARNING
drapes (such as textiles).
►
Risk of injury when twisting
the saw head in the ON posi-
tion through inadvertent acti-
WARNING
vation of the product!
►
Risk of injury caused by inad-
vertent loosening of the saw
blade during operation!
WARNING
►
►
Note
The motor found within this Acculan 3Ti hand piece is
powered by a magnetic system. To ensure that the
motor of this powered instrument does not inadver-
tently activate, do not allow the hand piece to come in
contact with any magnetic sources (i.e. magnetic
instrument pads). Please note that having the safety
switch in the locked position does not assure that the
motor will not run inadvertently.
Note
The low whistling noise at startup of oscillating saw 4 is
normal for this type and design of saw!
►
To run oscillating saw 4: Press oscillation frequency
control trigger 5.
The saw runs at the continuously controlled oscil-
lation frequency.
Do not let the tool come
into contact with drapes
(such as textiles) during
operation.
Only twist saw head in the
OFF position.
Do not press the button for
saw blade release/fixation
during operation.
Check for secure fixation
of the tool after any tool
change.