5. INSTALLATION
5.1 SETTING UP THE LIGHT SOURCE
Place the light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with
liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
DO NOT block the venting grids of the light source or impede
exhaust air flow. A minimum of 6"/15cm on each side and
12"/30cm at the back is required, 24"/60cm is preferred. The
lightsource should not be placed where its exhaust will
influence other devices, nor where the heated air from other
devices will influence the lightsource.
Make sure the rear power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
Make sure power cord is completely seated into the power inlet.
Use only cords provided with the light source.
CAUTION
Plug the AC power cord into a wall outlet using the three (3) prong plug supplied with the unit.
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power
with three prong plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
When used in clinical or residential areas near radio or TV units,
this equipment may be subjected to radio interference. To avoid
CAUTION
adverse electromagnetic effects, DO NOT operate this equipment
near RF energy equipment.
LIT124SUNOPTICS SURGICAL
(English)
®
1 2
3 4
Rev. G
Page 8 of 65