Chapter 11 Guide of EMC
The product is intended for use in the electromagnetic
environment specified below. The customer or the user
should assure that it is used in such an environment.
Guidance and manufacturer's declaration -
electromagnetic emissions
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
Class B
Harmonic
Not applicable
emissions
IEC 61000-3-2
Voltage
Not applicable
fluctuations/
flicker emissions
Guidance and manufacturer's declaration -
electromagnetic immunity
Immunity Test
IEC 60601-1-2
Test level
Electrostatic discharge
±8 kV contact
(ESD)
±2 kV, ±4 kV,
IEC 61000-4-2
±8 kV, ±15 kV air
Electrical fast transient
/burst
Not applicable
IEC 61000-4-4
Surge
Not applicable
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
Not applicable
power supply input lines
IEC 61000-4-11
Power frequency
30 A/m
magnetic field
50Hz/60Hz
IEC 61000-4-8
Conducted RF
Not applicable
IEC61000-4-6
Radiated RF
10 V/m
IEC61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mains voltage prior to application of the
test level.
Electromagnetic Environment -
Guidance
The product uses RF energy only for
its internal function. Therefore, the
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The product is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low voltage
power supply network that supplies
buildings used for domestic
purposes.
Compliance level
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
50Hz/60Hz
Not applicable
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
17
18