Página 2
Relative levels of latex allergen would be considered high at 100μ/g, medium at 10- 50μ/g, and below 5μ/g is considered very low or no activity. Dry natural rubber has a level of 1μ/g, therefore the Rhys Davies Exsanguinator should be considered as not being effected by the protein allergy.
Página 3
3.3. User Instructions Elevate the patient’s limb for 1 minute. (A) Roll the exsanguinator onto an operators arm until their hand shows from the opposite end to insertion. (B) Grasp the patients hand or foot, enclosing all digits. ...
Página 4
4.4. Disinfection (by hand) Suitable disinfectants are; quaternary ammonium compounds, isopropyl alcohol, chlorine bleach 0.5% and phenolic. CAUTION: Dilute all disinfectants in accordance with the manufacturer’s guidelines. Apply disinfectant by cloth, spray or disinfectant wipe. Following disinfection, wash off the product with clean warm water and dry thoroughly using a soft cloth.
Página 5
Documento n.° 992044-ES. Versión n.°6 – 30.03.23...
Página 6
2. Información sobre el producto Los sistemas de torniquete de Anetic Aid son conocidos por su fiabilidad y diseño accesible. Son muy populares en Reino Unido, puesto que son ideales para quirófanos, urgencias o unidades del dolor.
Página 7
3.3. Utilización Eleve la extremidad del paciente durante 1 minuto. (A) Introduzca el equipo de exanguinación en su propio brazo hasta que su mano asome por el extremo opuesto. (B) Agarre con firmeza la mano o el pie del paciente. ...
Página 8
4.3. Limpieza (a mano) Limpie el producto con ayuda de agua caliente y un detergente neutro. Seque la superficie por completo con un trapo suave. ADVERTENCIA: No sumerja el producto en agua u otro líquido. Es imprescindible secar bien el producto para su cuidado y mantenimiento. 4.4.