3.3. User Instructions
(A)
 Elevate the patient's limb for 1 minute.
 (A) Roll the exsanguinator onto an operators arm until their hand shows from the
opposite end to insertion.
 (B) Grasp the patients hand or foot, enclosing all digits.
 (C) Pull gently on the limb whilst rolling the exsanguinator, from operator to patient.
 (D) Roll the exsanguinator up to the level of the padded tourniquet cuff.
 NOTE: Adult legs may require assistance for rolling over the upper thigh.
 Once the tourniquet cuff has been inflated, the exsanguinator can be rolled off the
limb.
4.
Care & Maintenance
4.1. Storage
 Store in a cool dry place away from radiators and other heat sources (15 +/- 10°C).
 Store covered (in the polythene sleeve and cardboard box provided), away from
direct sunlight, ultra violet lighting systems and electrical equipment.
4.2. Inspection
 The product should be visually inspected for damage or deterioration on a daily
basis.
 The pressure should be checked periodically using a loosely inflated blood pressure
cuff rolled inside the exsanguinator, the sphygmomanometer will record the pressure
within the walls.
 The maximum inflated pressure should not exceed 60mmHg, corresponding to a
maximum external circumference of 47cm (18.5") of the device when new
(measuring tape provided).
 To inflate the device;
4.3. Cleaning (by hand)
 Clean the product with warm water and neutral detergent and dry the surfaces
thoroughly using a soft cloth.
Document No. 992044. Issue 6 – 24.10.22
CAUTION: Do not attempt to repair tears or splits.
 Use only the inflation kit provided.
 Locate the exsanguinator hidden valve, between the yellow moulded cross-
hairs, and gently insert the inflation kit needle.
 Connect a sphygmomanometer to the T-piece tubing (open ended).
 Inflate using the inflation kit hand bulb to 60mmHg, or if using the measuring
tape, until an external circumference of 46-47cm (18-18.5") is achieved. NOTE:
Following repeated use a larger circumference may be required to achieve the
stated internal pressure. Remove the inflation kit needle.
CAUTION: The device should not be over inflated.
CAUTIONS:
 Do not soak or immerse the product.
 Adequate drying is crucial in the care and maintenance of this product.
(B)
(C)
(D)