DMG MixStar-eMotion
The mixing speed can be increased or decreased by pressing the Plus but-
ton [8] or the Minus button [9], which is possible during extrusion as well.
1. Start mixing by pressing the Start/Stop button [5] or by operating the
foot pedal. If Timer 1 is set its countdown will start now. The end of
the working time is indicated by a long acoustic signal. If a new Mix-
Star cartridge is used discard the first 3 cm of the newly mixed mate-
rial. Carefully follow the manufacturer's instructions for the selected
material. For a clearer display all elements for which no new settings
are made (-:--) are hidden after the Start/Stop button [5] is pressed.
2. Tilt the tray slightly under the mixing tip to load it. To avoid entrap-
ment of air leave the mixing tip immersed in the impression material
while filling the tray continuously with the material.
3. After the tray has been sufficiently loaded stop dispensing by pressing
the Start/Stop button [5] or by releasing the foot pedal. Mixing is
stopped at once and pressure on the cartridge is relieved to avoid any
further flow of the material. If Timer 2 is set its countdown will start
now. 3 acoustic signals indicate when the material can be removed
from the mouth. The unit stops dispensing automatically after 2 min-
utes but can be restarted by pressing the Start/Stop button [5].
Filling an elastomer syringe
For filling elastomer syringes we recommend using mixing speed 1. The
mixing speed can be increased by pressing the Plus button [8] or decreased
by pressing the Minus [9] button. Speed adjustments are possible even
during material delivery.
1. Remove the plunger of the elastomer syringe.
2. Attach the elastomer syringe with its rear end to the mixing tip.
3. Start mixing by pressing the Start/Stop button [5] or by operating the
foot pedal. If Timer 1 is set its countdown will start now. The end of
the working time is indicated by a long acoustic signal.
Sofern Timer 1 eingestellt wurde, beginnt dieser nun seinen Count-
down. Das Ende der Verarbeitungszeit wird durch ein langes akustisches
Signal angezeigt.
4. Stop dispensing into the syringe by pressing the Start/Stop button [5]
or releasing the foot pedal. Mixing is stopped at once and pressure on
the cartridge is relieved to avoid any further flow of the material.
If Timer 2 is set its countdown will start now. 3 acoustic signals indicate
when the material can be removed from the mouth.
Exchanging a cartridge
1. The cartridge can only be removed if the plungers are in cartridge
exchange position. If necessary, press the Reverse button [6] and wait
until the plungers return to their original position. The Reverse indica-
tor [J] blinks during plunger retraction. When the plungers reach the
cartridge exchange position the Reverse indicator is no longer dis-
played.
2. Release the cartridge from the cartridge compartment by pressing with
one hand from below the openings and remove it from above. Leave
used mixing tip on the cartridge after use. It serves as a cap.
3. Insert the new cartridge into the unit from the top (see »Preparing the
cartridge for use«, step 5). In order to use up the material in the car-
tridge keep a new cartridge with attached mixing tip ready before start-
ing the mixing process. That way you ensure that the remaining mate-
rial in the old cartridge does not have to be discarded as the working
time limit would be exceeded while exchanging the cartridge.
Attaching a mixing tip
Attach a new mixing tip before each mixing process.
[Fig. 09 - 10]
1. Slide locking mechanism into open position.
2. Discard the old mixing tip. If the cartridge was stored for a longer peri-
od of time be sure to check the openings for possible blockings and
remove them.
3. Align the mixing tip with respective outlets on the cartridge and attach
it to the cartridge.
4. Lock the mixing tip with the locking mechanism on the cartridge.
Attach a new mixing tip before each mixing process. The device is
unable to deliver impression material through material that has cured
in the mixing tip. If the mixing tip is not properly attached to the car-
tridge material may be pressed past the mixing tip and cause soiling.
Setting the timer
The unit provides memory for 99 timer settings. The timers are used to
set the specific working time [B] and time in mouth [C] for the material
used. The maximum time setting is 9:59 minutes. An acoustic signal indi-
cates when the set time expires (0:00) and the timer is reset to the previ-
ous time setting. The previous time setting kept in memory even when
the unit is turned off. You assign your settings to a fixed program memo-
ry (D) and access them by recalling the respective program. The unit can
be operated with both, one, or none of the timers. The device is supplied
with preconfigured programs according to the DMG material specifica-
tions.
1. Press the Menu button [10] briefly. The program display is selected
and blinks.
2. Select a program memory to store the timer settings.
Use the Plus button [8] and Minus button [9] to scroll through the pro-
grams. Keeping either of the buttons pressed scrolls through the pro-
grams in steps of 10. Pressing one of the two buttons briefly opens one
setting at a time.
English
[Fig. 11 - 12]
[Fig. 03]