DMG MixStar-eMotion
Replace: Unscrew the screws of the plunger disks with the hex wrench
provided. Remove the plunger disks from their plungers and replace
them.
5. Press and hold the Menu button [10] again, and press the Reverse but-
ton [6]. The plungers return to their original position. Plunger disks and
hex wrench are included in the accessory »Service Kit«.
Replacing the fuses
1. Unplug power cable.
2. Take out the fuse drawer [2] above the power port on the back of the
device with a screw driver and replace the fuses.
Power input
Fuse type
Replacement fuses are included in the accessory »Service Kit«.
Troubleshooting
In case of a malfunction the device will display different error codes. E01
indicates a possible operation error. Please proceed as advised on the dis-
play. If the malfunction cannot be remedied or if another error code appears
please contact us
by phone at +49 40 84 00 60 or by email at info@dmg-dental.com.
Device does not start
• Check if the power cable is connected properly to the device port and
the outlet.
• Check the fuses and replace them if necessary. Follow the specifica-
tions on the rating plate.
Device starts but no impression material is delivered
• Check if a new mixing tip is attached.
• Foil bags: No foil bags in cartridge, foil bag is not opened sufficiently.
Impression material flows too slow
• Let impression material warm to room temperature (18 °C/64 °F).
• Foil bags: Check correct placement.
Impression material is dispensed, mixer does not turn
• Check if mixing rod is engaged properly into mixing tip.
• Attach a new mixing tip, if hex opening of mixing tip is damaged.
Damaged foil bags
• Check plunger disks. Replace damaged plunger disks, tighten plunger
disks if loose.
Device automatically switches off
• Remove cured material from cartridge openings.
• Check if a new mixing tip is attached.
• Switch off device with main switch for at least 5 seconds.
115 V
230 V
T 4 A, 250 V
T 2 A, 250 V
Packaging
1 MixStar-eMotion mixing device
230 V International
115 V International
115 V USA
230 V GB
230 V NZ/AUS
Accessories
1 foot pedal
50 MixStar-Tips
MixStar Service Kit
Plunger disks, fuses, Allen wrench
1 Service-Kit 115 V
1 Service-Kit 230 V
English
REF 900480
REF 900482
REF 999482
REF 900483
REF 900484
REF 900310
REF 909567
REF 900488
REF 900490
17