Use Method Of The Curing Stand Set - gc Labolight DUO Instrucciones De Uso

Unidad de fotopolimerización led dual para resinas de composite indirecta
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 118
2) A prosthesis or artifi cial tooth with a height of 52mm or less placed on a model
Use "Curing plate" and "Curing stand holder"
70mm directly
(Recommended curing position)
3) A prosthesis or artifi cial tooth with a height of 52mm or more placed on a model
Place it together with the model on the turntable.
● When placing an object subject to curing outside the recommended curing range, set curing time rather
larger since light may not suffi ciently shine on the object.
● Care should be taken that if a prosthesis or model placed outside the 100mm outer diameter of the
turntable or higher than 80mm, interference of the object subject to curing or model with the visor or the
curing chamber wall occurs, thereby causing improper or unable rotation of the turntable.

5.4 Use method of the Curing stand set

For a prosthesis alone, such as crown and bridge, or that placed on a model equipped with dowel pin, set it in
combination with the curing stand and various posts by referring to the following examples of use.
Clip
Fixes a prosthesis and inserted into a round hole or three
consecutive round holes provided on the curing stand.
Metal post
Receives a prosthesis and inserted into a round hole or three
consecutive round holes provided on the curing stand.
Silicone post
Receives a prosthesis and inserted into a hexagonal hole
provided on the curing stand. Cut the protrusion for use, as
needed.
Prosthesis placed on a model equipped with a dowel pin
The model with a prosthesis thereon is placed on the curing stand by inserting the dowel pin into the
hexagonal hole provided on the stand.
EN 20
70mm
Note
Curing stand
Three consecutive
round hole
52mm
Round hole
Hexagonal hole

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido