Amersham Pharmacia Biotech EPS 3501 XL Manual Del Usuario página 9

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

1. Choose Program Number
(1→2)
2. Choose Step or Gradient
mode. (Step)
3. Confirm Phase number (1)
4. Set Voltage limit (step) or
Voltage endpoint
(gradient).
(0→1000 V)
5. Set Current limit.
(0→10 mA)
6. Set Power limit
(0→10 W)
7. Choose Breakpoint unit
(h, Vh or mAh) or no
breakpoint (OFF).
(h→Vh)
(h→OFF)
8. Set Breakpoint
(0→1000 Vh)
9. Last Phase,
YES or NO?
10. Choose Alarm OFF or ON.
(ON)
11. The program is ready.
Fig. 3. Step-by-step summary of programming.
SET:
1 1
500 V
1 mA
Select by
Confirm by
SET:
2 1
STEP/gradient
Select by
Confirm by
SET:
2 1
0 V
0 mA
set
Confirm by
enter
SET:
2 1
0 V
0 mA
Select by
Confirm by
SET:
2 1
1000 V
10 mA
Select by
Confirm by
SET:
2 1
1000 V
10 mA
Select by
Confirm by
1st Phase
Other phases
SET:
2 1
1000 V
10 mA
Select by
Confirm by
Vh
SET:
2 1
1000 V
10 mA
Select by
Confirm by
SET:
2 1
Last phase
?
Select by
Confirm by
YES
Alarm Phase 1?
Phase: - - - -
Select by
Confirm by
A
SET:
2 1
1000 V
10 mA
1 Vh
5 W
set
enter
set
enter
0:00 h
0 W
0:00 h
0 W
set
enter
0:00 h
0 W
set
enter
0:00 h
0 W
set
enter
0:00 h
10 W
set
enter
0 Vh
10 W
set
enter
1000 Vh
Y
set
enter
NO
N
set
enter
1000 Vh
SET:
10 W
1000 V
Breakpoint OFF
2 1
OFF
10 mA
10 W
7