5 Using your STATIM
5.3.2. Wrapped Cycle
(STATIM 5000 G4)
The Wrapped Cycle is used to sterilize up to
1.5 kg (3.3 lbs) of solid and hollow metal
instruments which have been sealed in paper /
paper, or paper / plastic autoclave bags. Dental
hand pieces may be sterilized in this cycle.
To select the Wrapped Cycle, press the Wrapped
Cycle button, then press the START button.
The sterilization temperature in the cassette is 134˚ C (273˚ F) and the holding time is 6 minutes.
See Section 4.4 Preparing and Loading Instruments before running this cycle.
If wrapped instruments are intended for storage, the wraps must be dry when the cassette is removed
from the unit and opened.
5.3.3. Rubber and Plastics Cycle
(STATIM 5000 G4)
The Rubber and Plastics Cycle is used to sterilize
up to 0.4 kg (0.9 lbs) of solid or hollow unwrapped
instruments constructed of metal or the materials
listed in Section 4.4 Preparing and Loading
Instruments.
To select the Rubber and Plastics Cycle, press the
Rubber and Plastics Cycle button, then press the
START button.
The sterilization temperature in the cassette is 121 ˚C (250 ˚C) and for the holding time is 15 minutes.
A
A
B
C
A
A
B
C
Page 30
D
B
C
D
Warming up
E
Conditioning
F
Pressurizing
D
B
C
D
Sterilizing
Warming up
E
Conditioning
Venting
F
Air Drying
Pressurizing
E
F
Sterilizing
Venting
Air Drying
E
F

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Statim 5000 g4

Tabla de contenido