Guidance And Manufacturer's Declaration; Electromagnetic Emission; Electromagnetic Immunity - .bock eloflex Manual De Instrucciones

Tabla de contenido

Publicidad

Guidance and manufacturer's declaration

Electromagnetic emission

The medizinisches Bett is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the medizinisches Bett should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11 (partly)
RF emissions
CISPR 11 (partly)
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations/fl icker
emissions
IEC 61000-3-3

Electromagnetic immunity

The medizinisches Bett is intended for use in the electromagnetic environment specified below.
The customer or the user of the medizinisches Bett should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrostatic transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic fi eld
IEC 61000-4-8
NOTE: U
is the a. c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
T
(>95 % dip in U
) for 0,5 cycle
T
40 % U
T
(60 % dip in U
) for 5 cycles
T
70 % U
T
(30 % dip in U
) for 25 cycles
T
< 5 % UT
(>95 % dip in U
) for 5 sec
T
3 A/m
50
Electromagnetic environment - guidance
The medizinisches Bett uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
The medizinisches Bett is suitable for use in all establishments other than domestic and
those directly connected to the public-voltage power supply network that supplies buildings
used for domestic purposes.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply ines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
T
(>95 % dip in U
) for 0,5 cycle
T
40 % U
T
(60 % dip in U
) for 5 cycles
T
70 % U
T
(30 % dip in U
) for 25 cycles
T
< 5 % U
T
(>95 % dip in U
) for 5 sec
T
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile.
If fl oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the medizinisches Bett requires continued
operation during power mains interruptions, it is
recommended that the medizinisches Bett be
powered from an uninterruptible power supply
or a battery.
Power frequency magnetic fi elds should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.

Publicidad

Tabla de contenido
loading

Productos relacionados para .bock eloflex

Tabla de contenido