General Operation-Continued
4. Put the P2-E6 into the carry bag, position vents for exhaust away from user.
5. Connect the nasal cannula to the nozzle fitting.
CAUTION
Ensure that the cannula is routed to prevent it from being pinched or kinked to avoid a
disruption of oxygen flow.
CAUTION
The cannula is designed for disposable use.
6. Press On/Off
The device will beep, and the indicator light will flash.
"Welcome" will appear on the display at start up. The display will indicate the selected
flow setting and power condition. Two-minute warm up time will initiate. During this period
the oxygen concentration is building to the specified value but may not have yet reached
specification. Under special conditions, a longer warm-up time may be necessary, such as
in extremely cold temperatures where the unit was stored or is being operated.
CAUTION
Oxygen concentration may not reach specification during the two-minute warm up time.
CAUTION
30 seconds after startup, the P2-E6 will enter auto-pulse mode. During these 30 seconds,
inhalation will not work.
Operating Instructions
Nozzle fitting is located on the top side of the P2-E6 near the Pre-Filter
Cover. Connect a nasal cannula to the nozzle fitting on the device.
(pictured)
button to turn on P2-E6.
21