EMC Information Continued
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Radiated RF
IEC 61000-4-3
Guidance and Manufacturer's Declaration -
Electromagnetic Emissions:
The P2-E6 is intended for use in the electromagnetic environment specified below. The
user of the P2-E6 should ensure it is used in such an environment.
Emission Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC61000-3-2
Voltage
Fluctuations/Flicker
Emissions IEC61000-3-3
System Specifications
3 Vrms
150 kHz to 80 MHz
6 Vrms
Amateur Radio &ISM
Bands between 150 kHz
and 80 MHz
10 V/m
80 MHz to 2.7GHz
Compliance
Group 1
Class B
Class A
Complies
3 Vrms
150 kHz to 80
MHz
6 Vrms
Amateur Radio
&ISM Bands
between 150
kHz and 80
MHz
10 V/m
Electromagnetic Environment -
Guidance
The P2-E6 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
The P2-E6 is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
32
Portable and mobile RF
communications equipment
should not be used closer
(to any part of the device,
including cables) than the
recommended 11.8" 30 cm
separation distance.
Interference may occur in
the vicinity of
Equipment marked with the
fallowing symbol: