ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1 - Emission
Phenomenon
Compliance
CISPR 11
RF emissions
Group 1, Class B
IEC 61000-3-2
Harmonic distortion
Class A
IEC 61000-3-3
Voltage fluctuations and flicker
Compliance
Table 2 - Enclosure Port
Basic EMC
Phenomenon
standard
Electrostatic
IEC 61000-4-2
Discharge
Radiated RF EM field
IEC 61000-4-3
Proximity fields from RF
wireless communications
IEC 61000-4-3
equipment
Rated power frequency
IEC 61000-4-8
magnetic fields
Electromagnetic environment
Home healthcare environment
Home healthcare environment
Home healthcare environment
Immunity test levels
Home Healthcare Environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Refer to table 3
30A/m
50Hz or 60H
32
Table 3 – Proximity fields from RF wireless communications equipment
Test frequency
Band
(MHz)
(MHz)
385
380-390
450
430-470
710
745
704-787
780
810
870
800-960
930
1720
1845
1700-1990
1970
2450
2400-2570
5240
5500
5100-5800
5785
Immunity test levels
Professional healthcare facility environment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine, 28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
33