INT500
voltage will continue to be applied until the TEST/LOCK
button is pressed again.
4. Stop the output test voltage before removing the test leads
(to enable the tester to discharge capacitive circuits).
If the screen displays volts, wait until it reaches zero.
5. The meter displays the " > " symbol and the maximum
resistance for the range when measured resistance is
higher than the maximum display range.
Using the Compare function
Before starting the Insulation Resistance test, select the
comparative value in Setup mode: 0.5MΩ, 10MΩ, 20MΩ,
50MΩ, 100MΩ, 200MΩ, 500MΩ, 1000MΩ.
Before starting the Earth-Bond Resistance test, select the
comparative value in Setup mode: 0.5Ω, 1Ω, 2Ω, 3Ω, 4Ω, 5Ω,
10Ω, 20Ω, 30Ω, 40Ω.
If the measured value is better than the selected compare
value, the Pass indicator will be green, otherwise it is red.
100M
100M
V
Press
COMPARE
11
EN
100M
100M
M