tion/inhalation solutions) for high-risk groups
(e.g. patients with cystic fibrosis).
• Ensure thorough drying after each cleaning
and disinfection process. Residual moisture
or wetness can represent an increased risk of
bacterial growth.
Preparation
• Immediately after each treatment, all parts of
the atomiser
and the rest of the atomiser set
8
used must be cleaned of medication residues
and contamination.
• To do this, dismantle the atomiser
individual parts.
• Remove the mouthpiece
nosepiece
from the atomiser.
21
• Disassemble the nosepiece if you have used it
with the comfort attachment.
• Dismantle the atomiser by twisting the top an-
ticlockwise against the medication container
.
11
• Remove the atomiser insert
cation container
11
• Remove the valve
atomiser
.
8
• Reassembly is carried out in reverse order.
Cleaning
Important
The device must be switched off, disconnected
from the mains and allowed to cool down each
time before cleaning.
The atomiser and the used atomiser set, such as
the mouthpiece, masks, atomiser insert, nasal
douche, etc. must be washed with hot but not
boiling water after each use. Dry the parts careful-
ly using a soft cloth. Put the parts together again
when they are completely dry and place them in a
dry, sealed container or disinfect them.
When cleaning, ensure that any residue is re-
moved. Never use any substances for cleaning
that could potentially be toxic if they came into
contact with the skin or mucous membranes, or if
they were swallowed or inhaled.
Use a soft, dry cloth and non-abrasive cleaning
products to clean the device.
Do not use any abrasive cleaning products and
never submerge the device in water.
Important
• Make sure that no water gets inside the device!
into its
8
, mask
12
13 14
from the medi-
9
.
by pulling it out of the
10
• Do not clean the device or atomiser set in the
dishwasher!
• Do not touch the device with wet hands when
it is plugged in and do not allow water to spray
on the device. Only operate the device if it is
completely dry.
• Do not spray any liquid into the vents! If liquid
penetrates the device this could cause dam-
age to the electrics or other nebuliser parts and
lead to a malfunction.
Condensation, hose care
Condensation may form in the hose depending on
the ambient conditions. It is essential to remove
the moisture to prevent bacterial growth and en-
or
sure proper treatment. To do so, proceed as fol-
lows:
• Disconnect the hose
• The hose must remain in the nebuliser
• Operate the nebuliser until the moisture is re-
moved by the air passing through.
• In the event of heavy contamination, replace
the hose.
Disinfection
Please carefully observe the points below when
disinfecting your atomiser and the atomiser set.
We recommend disinfecting the individual parts
on a daily basis after the last usage as a minimum
measure.
• First, clean the atomiser and atomiser set as
described in the "Cleaning" section.
• Place the disassembled atomiser
douche (n/a), the mouthpiece
masks
13 14
piece in boiling water for 5 minutes.
• For disinfection immerse the compressed air
hose in a 2 % sodium hypochlorite (NaOCl)
solution for 15 minutes (i.e. a solution made
from the disinfecting agent Amuchina
2  % sodium hypochlorite solution mixed by
your pharmacist). Afterwards rinse all parts
thoroughly with tap water, to fully remove any
possible residue of the disinfectant solution.
23
from the atomiser
7
, the nasal
8
, the silicone
12
and the disassembled nose-
.
8
.
4
®
or a