4. With dark grey fabric side facing inwards, insert both upper headgear
straps into the frame from the inside and fold the fastening tabs over
to secure.
Reprocessing the mask between patients
Only Sleep Lab Mask variants of the mask system are intended for multi-
patient re-use. This variant is identified by the "SLM" acronym on the
packaging label and a Unique Device Identifier labeling on the mask.
When using between patients, these masks must be reprocessed
according to instructions available on ResMed.com/downloads/masks.
Technical specifications
Mask setting options: For AirSense
Compatible devices: For a full list of compatible devices for this mask, see the
Mask/Device Compatibility List at ResMed.com/downloads/masks.
SmartStart may not operate effectively when using this mask with some CPAP or bi-level
devices.
Pressure-flow curve
Therapy pressure: 4 to 30 cmH
Resistance with Anti-Asphyxia Valve closed to atmosphere
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cmH
at 100 L/min: 1.7 cmH
Resistance with Anti-Asphyxia Valve open to atmosphere
Inspiration at 50 L/min: 1.1 cmH
Expiration at 50 L/min: 1.1 cmH
Anti-Asphyxia Valve pressure
open-to-atmosphere (activation): <4 cmH
closed-to-atmosphere (deactivation): <4 cmH
™,
Mask Pressure (cmH
O
2
O
2
O
2
O
2
O
2
AirCurve
, and S9
O)
2
O
2
O
2
devices, select 'Pillows'.
Pressure
(cm H
O)
2
4
11
17
24
30
English
Flow
(L/min)
20
35
44
54
61
9