APPLICATIONS FOR DIFFERENT SPECIALTIES
1. General Dentistry
A. Assists in detection of caries. This technique is especially helpful for location of caries
which may be blocked on x-ray due to existing fillings.
B. Locates supra and sub gingival calculus.
C. When composite removal is required, locates residual composite material which appears
darker than surrounding tooth structure.
2. Prosthodontics
A. Location of the dentino-enamel junction in crown preparations.
B. Location of fractures in teeth and porcelain jackets.
3. Endodontics
A. Location of the root canal orifice.
B. Detection of broken instruments in the root canal.
C. Location of root fractures.
4. Periodontics and Dental Hygiene
A. Useful as an auxiliary light source during periodontal surgical procedures.
B. Assists in detection of sub gingival calculus.
C. Helps demonstrate decay and calculus directly in the patient's mouth and is a valuable aid
in patient motivation.
D. Oral hygiene instruction and routine screening procedures by the hygienist or assistant.
5. Orthodontics
A. Location of proximal caries when teeth are banded.
6. Pedodontics
A. Detection of crown fractures in traumatic injuries.
CLEANING/ DISINFECTING/ STERILIZING
a) Cleaning:
• Immerse/thoroughly drench parts in a surface decontaminate cleaner (soap and water cleaning solution).
• After 10 minutes, scrub with an instrument cleaning brush until visibly free from debris.
• Rinse all parts with warm water.
b) Sterilizing:
• Sterilize in a steam autoclave using distilled water at a temperature of 270°F (132°C)
for 20 minutes.
• Do not add chemical solutions to the autoclave.
• Do not use a chemical autoclave.
c) Disinfecting: The Microlux 2 LED hand piece and Endo-lite, Perio Probe and Proximal Carries
plastic fibers can be disinfected following these guidelines:
• Wipe all surfaces with a disinfectant comprising 72% - 55% alcohol and .33% -.5% quaternary ammo-
nium and follow the manufacturer's directions. Some popular disinfectant wipes in this category are:
1-Monarch Surface Disinfectant Wipes (Air Techniques)
2-Opticide 3 Wipes (Biotrol)
3-Super Sani-Cloth Wipes (PDI)
• Wipe, do not spray, solution onto the Microlux 2 hand piece. Prevent liquids from entering open-
ings on unit
MAINTENANCE
!
Batteries: When replacing batteries, it is important that they be placed in proper
polarity. Unscrew the End Cap and insert two AAA batteries with the positive (+) ter-
minal in the front towards the LED. Replace the End Cap. If the Microlux 2 is not going
to be used for more than 30 days, it is recommended that the batteries be removed. In order to
save battery life and to avoid leakage you may disconnect the battery by rotating the end cap
counter-clockwise by one full turn until you see the red O-ring.