ANALYSER COMPONENTS
The MyCare Insite consists of the MyCare Insite Analyser laboratory photometer and the MyCare Insite Touch
Screen. The Insite package contains:
•
User Manual
•
MyCare Insite Analyser
•
Power adapter
Before first use remove the transport lock from the door of the Analyser.
Note: Keep both the transport lock and the original packaging. In case of a defect, the device must
be returned in the original packaging with the transport lock installed to prevent damage in transport.
The Insite Technical Guide and videos for unpacking and setting up the Insite are at MyCareInsite.com.
With WiFi Connectivity tap
WARNINGS AND PRECAUTIONS
• For in vitro Diagnostic Use Only
• The Analyser performs self checks during measurements. Any electro-mechanical error will stop a result
from being generated. If an error message appears on the screen, please consult the "ERROR MESSAGES
AND POSSIBLE CAUSES" on page 10. Repeat the test.
• The door prevents ambient light, dust, dirt, and humidity from entering the Analyser during the testing process.
• Remove the test cartridge from the Analyser after every test. Dispose of the test cartridge appropriately.
• Make sure the door is closed whenever the Analyser is not in use.
• Do not attempt to open the Analyser door manually.
• Use only your fingertips to operate the Touch Screen! Do not use pens or other objects that may scratch or
damage the screen.
• Do not use liquid cleaners on the Analyser or Touch Screen. Use a lint-free, slightly damp cloth only.
• Connect optional equipment only when the Analyser is unplugged. Please note that attaching optional
equipment (e.g., a printer) can increase the amount of leakage current.
• Use ONLY manufacturer-approved tests. Otherwise, serious damage to the Analyser and/or inaccurate test
results may occur.
• To ensure samples are correctly identified run only one sample at a time.
• To ensure safe operation, always use the MyCare Insite according to the instructions in the package insert.
• Avoid electromagnetic radiation and electrostatic discharge.
• Prevent physical access by unauthorized users
• Enable screen lock setting "Password" on the Touch Screen.
• Any serious incident that occurs in relation to the MyCare Insite shall be reported to the manufacturer and
the competent authority of the Member State in which the user and/or the patient is established.
• Residual Risks: User error may cause display of a wrong measurement result, delayed result, no result; and
unauthorized access to information.
PI MCI-EUR-ML-PKG Rev 01
•
Power cable
•
Test Rack
•
USB cable
on the Main Menu to access the information.
•
Touch Screen
•
Touch Screen Stand
•
Touch Screen Mounting instructions