Loosen the 2 Allen screws on the lower sup-
port tube
Feed the support tube cable through the roll
tripod
Fasten the support tube to the roll tripod using
2 Allen screws and serrated washers (10 Nm)
The cable holder must be aligned at the rear
Connect the same colored stranded wires on
the lower support tube and the power supply
Further tighten down the power supply pan
with crosstip screws and serrated washers.
Perform this assembly step on an even surface:
3
2
metric
Loosen the grub screw 1
Place the spring balanced arm 2 on the upper
support tube
1
3
3
metric
Screw in the grub screw in the alignment
shown
In the case of a grub screw dead stop, screw
this back by a ¼ turn and check the turning
function
Push together the plug on the upper and lower
support tube until they click
Place the upper support tube on the lower one
3
metric
Tighten the cable holder using a 3-type Allen
screw on the upper support tube (2.4 Nm)
WARNING
Never unscrew the upper cable holder - risk of
injury.
If both cable holders are unscrewed, the
connecting piece will come loose and falls
down, which can lead to personal injury and
damage to the cable and device.
Continue with luminaire head assembly, see
Chap.7