OPERATIONAL STEPS AT THE END OF THE PROCEDURE
NOTE: ALWAYS USE INDIVIDUAL PROTECTION DEVICES (EX. LATTICE GLOVES) EVERY TIME YOU MAY GET IN CONTACT WITH POTENTIALLY
CONTAMINATED PARTS.
1.
DISCONNECT DISPOSABLE PARTS AND DISPOSE OF THEM.
2.
CHECK THE GENERAL INTEGRITY OF THE ASPIRATOR
3.
CARRY OUT THE FOLLOWING CLEANING OPERATIONS
CAUTION - DISPOSABLE DEVICE
DO NOT REUSE - RISK OF CROSS-INFECTION AND DETERIORATION
ASPIRATION TUBES SHOULD BE STERILIZED AFTER EACH USE.
THE FILTER ON THE APPLIANCE IS DISPOSABLE; THEREFORE IT IS NECESSARY TO REPLACE IT AFTER EVERY USE. DO NOT USE THE ASPIRATOR
WITHOUT A FILTER
JAR CLEANING
THE FLUID COLLECTION JAR CAN BE AUTOCLAVED AT A TEMPERATURE NOT EXCEEDING 121 ºC FOR NO MORE THAN MAX 3 MINUTES. IF
EXCEEDED, THE JAR MAY GET DAMAGED. OTHERWISE, THE JAR CAN BE DISINFECTED WITH DENATURED ALCOHOL OR A HYPO CHLORITE BASED
SOLUTION THAT CAN BE EASILY FOUND IN PHARMACIES. MAXIMUM REUSE LIMIT: 30 CYCLES
HOW TO CLEAN THE APPLIANCE
CLEAN IT WITH A SOFT CLOTH WETTED WITH NON ABRASIVE DETERGENTS
PERIODIC MAINTENANCE CHECK
ASPIRATOR F-170 WAS DESIGNED TO KEEP MAINTENANCE TO A MINIMUM. NEVERTHELESS IT IS NECESSARY TO CARRY OUT A DAILY CHECK
BEFORE USING THE APPLIANCE.
BEFORE CARRYING OUT ANY CHECK DUE TO MALFUNCTIONING OR ANOMALIES, CALL THE FAZZINI S.R.L. TECHNICAL SERVICE OR PERSONNEL
QUALIFIED BY FAZZINI S.R.L.
FAZZINI S.R.L. DOES NOT OFFER ANY WARRANTY ON APPLIANCE THAT UPON INSPECTION BY THE TECHNICAL SERVICE SHOWS SIGNS OF
TAMPERING.
THE OPERATOR MUST EXCLUSIVELY REPLACE THE SUPPLIED ACCESSORIES IF NECESSARY IN LINE WITH THE APPLIANCE'S USE INTENSITY
ASPIRATION TUBES SHOULD BE STERILIZED AFTER EACH USE. IN CASE OF REPAIR, THE USER SHOULD CALL FAZZINI S.R.L.
TRANSPORTATION COSTS MUST ALWAYS BE AGREED WITH FAZZINI S.R.L. BEFORE SHIPPING FAILED PIECES.
SHOULD ANY FLUID LEAK IN THE APPLIANCE, CALL THE TECHNICAL SERVICE OF FAZZINI S.R.L. FOR MAINTENANCE.
DO NOT OPEN THE UNIT.
STORAGE
THE ASPIRATOR MUST BE STORED IN A DRY LOCATION, PROTECTED FROM ATMOSPHERIC AGENTS
STORAGE TEMPERATURE: BETWEEN -10° AND + 55°
RELATIVE HUMIDITY BETWEEN 20 AND 80%
DISPOSAL
WITH REFERENCE TO THE DISPOSAL OF THE APPLIANCE, ITS ACCESSORIES AND ITS SINGLE USE DISPOSABLE PARTS, PLEASE SEE CURRENT WASTE
DISPOSAL REGULATIONS
DO NOT ACT DIRECTLY ON THE APPLIANCE
FAZZINI S.R.L. RESERVES ALL THE RIGHTS TO CHANGE THIS HANDBOOK WITH NO PRIOR NOTICE FURTHER TO MANUFACTURING CHANGES TO
THE APPLIANCE, WHICH HAS BEEN BUILT IN COMPLIANCE WITH THE FOLLOWING STANDARDS:
UNI EN ISO 10079/2:2000
DEVICE IN CONFORMITY WITH 93/42 CEE -- 2007/47CE
FAZZINI S.R.L . IS COMMITTED TO THE MONITORING OF THE MEDICAL APPLIANCES IT DISTRIBUTES ON THE MARKET, AND THEREFORE SHALL BE
MOST GRATEFUL OF BEING INFORMED OF ANY PROBLEMS OR SUGGESTIONS RELATING TO ITS PRODUCTS, TO BE SENT IN WRITING TO:
FAX 0039/0227409242
E-MAIL FAZZINI@FAZZINI.IT
PLEASE STATE YOUR CONSENSUS TO DATA PROCESSING UNDER LAW 196/03
User Manual_F-170_Rev.5_16_09_2021
Rev.5 del 16.09.2021
Istruzioni per l'uso
Instructions for use
F-170
Mode d'emploi
Manual de instrucciones
Instruções de uso
Pag. 6