1. Presentation
This medical device connects to a medical gas wall outlet to supply another medical device with a medical gas or a mixture of medical gases.
The low-pressure hose is fitted with inlet and outlet connectors (or "terminal units") specific to each gas, in accordance with national
standards.
User profile:
Users are medical staff: doctors, nurses.
2. Versions
- Length of the hose: 0.5 m to 7 m.
- Connectors: comply with standards
- Compatible gases: Air, O
2
3. Technical data
- Complies with standard ISO 5359: 2014.
- Anti-static PVC hose for O
- Inside diameter of hose: 6.3 ± 0.3 mm.
- Outside diameter of hose: 12.4 ± 0.3 mm.
- Maximum operating pressure: 15 bar.
- Burst pressure: 60 bar.
- Ambient temperatures for use: - 15°C to +60°C.
- NON-sterile when supplied
- Does not contain latex.
4. Instructions for use
Verification before bringing into service: dropping the flexible hose may damage the inlet or outlet connector and make it impossible to
connect up the hose. Always ensure that the flexible hoses are checked before they are brought into service.
Connection: Start by connecting the hose assembly to the medical device it is intended to supply, then connect it to the medical gas wall
outlet.
Disconnection: Start by disconnecting the hose assembly rom the wall outlet, then disconnect the connector from the medical device.
5. Symbols
MANUFACTURER
6. Safety advice
- Do not connect two flexible hoses together in series (i.e. in a line) since this may reduce the pressure and flow.
- To prevent any risk of injury resulting from a release of pressure, hold the connector firmly when disconnecting.
- To prevent any risk of fire, explosion or intoxication:
o Do not operate the device near a source of ignition.
o Do not expose the hose to intense heat or high temperatures.
o Do not allow the device to come into contact with greasy or oily products.
o Do not modify the terminal units or threaded connectors crimped onto the flexible hose.
- The flexile hoses may only be used by personnel who have received training in the correct use of medical gases.
- DIN and Nordic 90-degree elbow connectors for O
7. Cleaning
- Use a disinfectant cleaner for medical devices. Leave to dry before using the device again.
- Check the compatibility of the disinfectant cleaner with the materials used to make the flexible hose (PVC and PP).
- Do not spray disinfectant cleaner directly onto the flexible hose. Use a cloth or a disposable wipe.
- Do not immerse the flexible hose in water or in any other liquid.
8. Storage
- Store at between -20 and +60°C in a clean and dry place.
- Protect from ultraviolet light.
- Keep the instructions with the packaging for as long as possible.
9. Maintenance
- The device must be inspected once a year (as a minimum).
- Check that the flexible hose is airtight: to test the airtightness, pressurise the flexible hose. Apply a leak detector to the flexible hose and
to the connectors. Check that no bubbles appear. After checking, clean the flexible hose thoroughly as described above.
- Check the connectors: check the condition of the connectors and that it is not possible to connect the hose assembly to a supply of gas
other than the gas for which the hose is intended.
In the event of damage or malfunction, return the device to the manufacturer or to an approved repairer.
10. Guarantee
- The device is covered by a 1-year warranty (parts and labour), which does not cover damage or incidents caused by negligence,
misuse, incorrect supervision or maintenance.
- Service life: 10 years.
- Date when
marking first affixed: 2000
TECHNOLOGIE MEDICALE must be notified immediately of any incident or near-incident
NF S 90-116 -BS 5682 -DIN 13260 - US Ohmeda Diamond - NORDIC - DISS - NIST - UNI - CARBUROS
, N
O, Vacuum.
2
, Air, N
O.
2
2
XXXX-XX
and Air are fitted with a non-return valve.
2
In accordance with European Directive 93/42/EEC of June 14th 1993,
- Does not contain phthalates.
- Connectors are crimped irremovably and permanently to the hose.
- The Lot No. specific to the hose assembly is inkjet printed onto
the device.
- Information engraved onto the crimping sleeve: year and month of
manufacture of the flexible hose + "TM CE 0459".
- Label on packaging indicates expiry date and the manufacturing
lot No. for the flexible hose.
USE UNTIL END YYYY-MM