When using in a place where static electricity is likely
to be generated or in a wet place, join the grounding
wire to the protective grounding terminal (screw).
6.
・Refer to "2.Cautions" regarding use.
・Refer to
"6.2.Error display
・Refer to the [FG-CWP Series Instruction Manual]
(https://www.aandd.jp/products/weighing/scale/platform/fgcwp.html
https://www.aandd.jp
・Periodically check the accuracy of weighing. Adjust the scale, if it is moved to another location or the
environment has changed.
6.1. Repair
Do not disassemble / assemble the scale without an authorized service engineer. Doing so may
cause an electric shock or damage to the scale, etc. In this case, repair is not covered under
warranty. Contact your local A&D dealer if your scale needs service or repair.
6.2. Error display
In this situation
Nothing displayed.
Scale does not turn on.
The scale does not
display zero at first.
e is displayed.
lo ut is displayed.
-Cal e is displayed.
lb is displayed.
p------- is
displayed and does not
proceed.
Maintenance
" and corresponding mode for displayed error code.
for sensitivity adjustment.
・Is the main power the correct voltage?
・Is the main power cord properly connected?
・Check around the weighing pan.
・Is there anything on the weighing pan?
・Refer to the [FG-CWP Series Instruction Manual]
(https://www.aandd.jp/products/weighing/scale/platform/fgcwp.html
)on our website
sensitivity adjustment.
・Weighing error that meaning "Over loaded."
・Indicates that the sample weight is too light to set the unit weight in
the counting mode.
・Sensitivity adjustment error that meaning "Too light."
・Displayed when the main power drops.
・The weighed value is unstable due to drift, vibration or other
factors.
A breeze or vibration may be affecting measurement.
Check around the weighing pan.
・Check the connection of the load cell cable.
・No zero display when the display is turned on.
Remove anything that is on the pan.
・Refer to the [FG-CWP Instruction Manual]
(https://www.aandd.jp/products/weighing/scale/platform/fgcwp.html
on our website (https://www.aandd.jp)and perform zero point
sensitivity adjustment.
Protective
Grounding
Terminal
Confirm these items
https://www.aandd.jp
8
Rear of Display Unit
)our website
and perform zero point
)