C. Connector
D. Pumping device
E. Suction bowl
F. Valve x2
G. Milk bottle
H. Bottle holder
I. Bottle teat
J. Charging cable
K. Bottle cap
03. DESCRIPTION OF MODES
Stimulation mode:
L1 - 5 kPa; L2 - 12 kPa; L3 - 17 kPa; L4 - 21 kPa ; L5 - 25 kPa ;
Expression mode:
L1 - 21 kPa; L2 - 24 kPa; L3 - 27 kPa; L4 - 31 kPa ; L5 - 37 kPa ; L6 - 40 kPa ; L7 - 42
kPa; L8 - 43 kPa; L9 - 45 kPa;
Allowable difference is ± 5 kpa
04. DESCRIPTION OF ICONS AND BUTTONS
SEE FIG. B
A. Suction decrease
B. Stimulation mode
C. Battery Percent
D. Power button/ mode changeZwiększenie siły odsysania
E. Suction increase
F. Suction level
G. Expression mode
H. Session Duration
1. Power button:
Hold for 2 seconds to turn the breast pump on or off.
2. Mode change:
Press the button to change the mode. The device will automatically switch from
stimulation to expression mode after 2 minutes. The device can save settings
and has a time counter. When you
turn on the device, the settings from the previous use of the device are
automatically turned on. The breast pump will automatically turn off after 30
minutes of use.
3. Suction power:
+ increases suction power; - reduction of suction power (stimulation mode:
L1-L5; expression mode: L1-L9).
4. Battery status:
Indicates the amount of battery power remaining. Charge the breast pump if
the battery is low.
12