7.0 • MAINTENANCE
Please contact your supplier if you require assistance with the maintenance of the rollator. Incorrect adjustments can result
in serious injury.
1.
Check all button clips are fully clicked in position and that both hand wheels are tight and secure.
2.
Check the wheels can rotate freely.
3.
Check for general wear or damage. Any suspected faults or damage should immediately be reported to the supplier and the
product taken out of use.
4.
Make sure the brakes are functioning correctly. When the brake levers are squeezed, the two rear wheels should not be able to
rotate.
7.1 • Adjusting the Brakes :
The following section requires some technical ability, do not attempt to carry
out the adjustment unless you are confident to do so.
1.
Refer to Figure 11.
The brake adjustment nut is located at the bottom of the brake cable,
above the rear wheel on both sides.
2.
Use a 10 mm spanner to loosen the locking nut in an anti-clockwise
direction.
3.
Use an 8 mm spanner to adjust the brake tension. To tighten the brakes,
turn the adjustment nut in an anti-clockwise direction.
Note:
There
must
mechanism and the wheel, when the brake lever is released.
Use a 10 mm spanner to tighten the locking nut in a clockwise direction.
Note : use the 8mm spanner to make sure the adjustment nut remains in
position, While tightening the locking nut.
Test the brakes in accordance with Section 5.1.
4.
Note : Incorrectly adjusted brakes can lead to injury, please contact head
office if you require assistance.
8.0 • PRODUCT DISPOSAL
This product should be thoroughly cleaned prior to disposal.
This product is manufactured from metal and plastic and can be disposed of through Local Authority Household Waste and Recycling
Centres or through Local Trade Waste Collectors.
9.0 • GARANTIE
The warranty period of the product is 2 years. It covers defects in workmanship, materials and hidden defects.
The warranty excludes damage related to structural modifications to the walker, insufficient maintenance, improper handling or storage,
or the use of parts other than the original parts.
In addition, the warranty excludes parts subject to natural wear.
Importer :
VILGO
Route de Sainte Alvère - Creysse BP212
24102 BERGERAC CEDEX
Tel : +33 (0)5 53 74 45 50
Fax : +33 (0)5 53 63 06 07
be
a
small
gap
between
FOSHAN OSCAR MEDICAL INSTRUMENT
ADD : Part 2, warehouse C, Nanhai
National Eco-industrial park, Danzao town,
Nanhai district, Foshan city
the
brake
FIGURE 11
Maker :
CO. LTD
7
Adjustment Nut
8 mm
Locking Nut
10 mm
Brake
Distributor :
Oxypharm
39 rue des Augustins
76178 Rouen Cedex