SPECIFICATIONS
Catalog Number
Delivery Rate
Maximum Recommended Flow (@
nominal outlet pressures of zero and
7 kPa)**
Outlet Pressure
Auxiliary Oxygen Port **
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 70˚F {21˚C} only)
0-1500 M (0-5000 ft)
Operating Environment Range*
41˚F (5˚C) to 95˚F (35˚C), humidity
range of 15% to 93% non-
condensing
Power Consumption
Weight
Safe Working Load
Sound Pressure Level at 3 and 10LPM
Sound Power Level at 3 and 10 LPM
Dimensions
Maximum Limited Pressure
Operating System
The visible "low oxygen" indicator
will activate at the following level
Storage Conditions
Equipment Class and Type
Approval Body and Safety Standard
CE mark
EMC Compliance To
*NOTE– The OSD performance at 41˚F (5˚C) to 95˚F (35˚C), 93% R.H. through voltage range on the 1025 verified at 670m.
**The maximum recommended flow is 6 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
Specifications subject to change without notice.
EN - 10
DEVILBISS 10-LITER SERIES
1025DS
2 to 10 LPM
10 LPM
20 .0 ± 1 .0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
120 V, 60 Hz, 6 .1 Amp
102-132 V~, 60 Hz
2-10 LPM=87%-96%
Across the voltage range:
No degradation of performance
No degradation in performance across
the operating voltage range .
120 vac, 60HZ: 639 watts average
42 lbs . (19 Kilograms)
53 lbs . (24 Kilograms)
<59 dBA
<69 dBA
24 .5"H x 13 .5"W x 12"D
(62 .2 x 34 .2 x 30 .4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
86% ± 3%
(The audible alarm will alert at
approximately 85% . At less than 60%, the
red "service required" light will activate .)
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing,
including condensation
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IP21
TUV
ANSI/AAMI ES60601-1:2005+A2 (R2012) +A1
IEC 60601-1-6:2010
IEC 60601-1-11:2015
ISO 80601-2-69:2014
CAN/CSA-C22 .2 No . 60601-1:14
CAN/CSA-C22 .2 No . 60601-1-6:11
CAN/CSA-C22 .2 No . 60601-1-11:15
CAN/CSA-C22 .2 No . 80601-2-69:16
No
EN60601-1-2
1025KS
2 to 10 LPM
10 LPM
20 .0 ± 1 .0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
230 V~, 50 Hz, 3 .2 Amp
195-253 V~, 50 Hz
2-10 LPM=93% +3%/-6%
Across the voltage range:
No degradation of performance
No degradation in performance across
the operating voltage range .
230 vac, 50Hz: 664 watts average
42 lbs . (19 Kilograms)
53 lbs . (24 Kilograms)
<57 dBA
<67 dBA
24 .5"H x 13 .5"W x 12"D
(62 .2 x 34 .2 x 30 .4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
86% ± 3%
(The audible alarm will alert at
approximately 85% . At less than 60%, the
red "service required" light will activate .)
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing,
including condensation
Class II Equipment Double Insulated;
Type B Applied Part
IP21
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
Yes
EN60601-1-2
1025UK
2 to 10 LPM
10 LPM
20 .0 ± 1 .0 psi (138 kPa +/- 7 kPa)
Outlet Pressure: <15 psi
Outlet Flow: 2 LPM
240 V~, 50 Hz, 3 .2 Amp
204-264 V~, 50 Hz
2-10 LPM=93% +3%/-6%
Across the voltage range:
No degradation of performance
No degradation in performance across
the operating voltage range .
240 vac, 50Hz: 670 watts average
42 lbs . (19 Kilograms)
53 lbs . (24 Kilograms)
<57 dBA
<67 dBA
24 .5"H x 13 .5"W x 12"D
(62 .2 x 34 .2 x 30 .4 cm)
35 PSIG (241 kPa)
Time Cycle / Pressure Swing
86% ± 3%
(The audible alarm will alert at
approximately 85% . At less than 60%, the
red "service required" light will activate .)
-13˚F (-25˚C) to 158˚F (70˚C), humidity
range of 15% to 93% non-condensing,
including condensation
Class II Equipment Double Insulated;
Type B Applied Part
IP21
TUV
IEC 60601-1:2012
IEC 60601-1-6:2010+A1
IEC 60601-1-11:2015
EN ISO 80601-2-69:2014
Yes
EN60601-1-2
A-1025