Reconditioning instruction/Re-use
en
- Alto can be re-used.
- Maintenance, inspection, and cleaning must be carried out regularly.
- When replacing parts, use original parts from Etac.
- When installing spare parts, the assembly instructions must be
followed. The assembly instructions and spare parts list can be found
at www.etac.com.
Be observant of malfunctions, mechanical wear, and cracks.
Check and test the following:
1
- That screws for the castors are properly tightened.
- That the locking function works. It should not be possible
for the castors to roll when locked.
- That the castors rotate.
2
Actuator, control box and hand control
- Cables and connections.
- Covers.
- The actuator's fixing points.
- Test the function of the hand control by tilting and raising/lowe-
ring the seat to its end positions.
- Test the function of the emergency lowering button.
Charge the batteries at least every 12 months.
Cleaning must be carried out regularly. It may be necessary to clean the
device between users to avoid infection.
Clean the device and its accessories in a washing machine or washing
tunnel (unless otherwise specified).
Cleaning by washing machine or washing tunnel must be carried out in
accordance with EN/ISO 60601-2-52.
The following points must be observed during the washing process:
- The device must be secured and positioned upright with the
wheels locked.
- In order to maintain the flexibility of the cables, it is important that
5-year warranty against material and manufacturing defects (1-year
warranty for the battery). For terms and conditions, see www.etac.com.
Castors
Lithium-ion battery
Cleaning and disinfection
Cleaning in a clinical environment
General
- Remember that the service life of the device is limited (see manual).
This is particularly important when instructing users who are expected
to use the product for an extended period of time.
Maintenance
3
- That the device and any accessories are assembled correctly
according to the assembly instructions
- That all screws are tightened
4
- That arm supports and arm support attachments are intact
- That the arm supports can be folded up/down
5
1. Do not detach any cables unless necessary.
If cables are disconnected the O-rings must be replaced.
O-rings and sockets must be lubricated with acid-free Vaseline.
2. Connect the electrical components according to the
wiring diagram.
3. Plug the mains cable into the wall socket for charging.
4. Perform calibration/reset; see the Troubleshooting guide
section in the user manual.
Clean the device using a standard non-abrasive cleaning agent
(pH between 5 and 9). Rinse and dry.
Disinfect the device with a 70% disinfectant solution. Allow to air dry.
the cable is positioned in such a way that its own weight does not load
the spiraling.
- The cables must be connected to prevent water ingress.
- To ensure that the pre-lubricated inner tube remains lubricated, the
actuator may only be washed when the piston rod is fully retracted.
When washing the soft back, the velcro straps must be
put together.
Miscellaneous
White touch-up paint can be purchased from Etac to touch-up minor
chips. Touch-up paint white RAL 9003. Art. no. 84005052
4