Instructions for Use
Lock Knob until the red line disappears, indicating the knob is tight.
NOTE: This clamp is equipped with a safety clutch. If the Rotation Knob is
securely tightened but the internal locking teeth are not fully
engaged, they will "pop" into position when loaded. If you hear
the "pop" when applying a load to the accessory, relieve the
load and securely retighten the Rotation Knob.
NOTE: To obtain maximum service life from this clamp always completely
unlock the Rotation Knob by turning it 2 full turns prior to adjusting
the angle of inclination.
Equipment removal:
a. Remove any load from the accessory prior to removing the socket.
b. Loosen the Power Grip Lock Handle and remove the accessory post from the
socket.
c. Unscrew the Rail Lock Knob completely.
d. Lift and remove the socket from the rail.
Cleaning and Maintenance
WARNING:
•
Do not use bleach or products that contain bleach to clean the device. Injury or
equipment damage can occur.
•
Clean and disinfect using a quaternary ammonium disinfecting/cleaning solution
following the manufacturer's recommendation for achieving low-level disinfection.
•
Do not immerse
•
After each use, clean the device with alcohol-based wipes.
•
Do not put the device into water. Equipment damage can occur.
•
Use a cloth and a quaternary ammonium disinfecting/cleaning solution to clean
and disinfect the device.
•
Read and follow the manufacturer's recommendation for low-level disinfection.
•
Read and follow the cleaning product's instructions. Use caution in areas where
liquid can get into the mechanism.
•
Wipe the device with a clean, dry cloth.
•
Make sure that the device is dry before you store it or use it again.
CAUTION: Damage may result if the product is cleaned with bleach.
©2019 Allen Medical Systems, Inc. All Rights Reserved
Document Number: 80028552
Version: A
Page 7
Ref Blank Template: 80026386 Ver. B
Issue Date: 19 MAY 2020