General specifications
Specification
Dimensions (W x H x D)
Sample contact components Stainless Steel, UHMW, Epoxy
Weight
Pollution degree
Installation category
Protection class
Power requirement options
Operating temperature
Maximum relative humidity
Storage temperature
Speed of response
Sample flow rate
Sample temperature
Plumbing connection
Communications
Outputs (optional)
Digital inputs (optional)
Certifications
Warranty
Details
320 x 750 x 250 mm (12.6 x 29.5 x 10 in.)
21.5 kg (47.4 lb) standard option
2
II
I
230 VAC, 50 Hz at 1.6 A ; 115 VAC, 50 Hz at 3.2 A
120 VAC 60 Hz at 3.2 A
0 to 50 °C (32 to 122 °F)
95% for temperatures up to 50 °C (122 °F)
-20 to 70 °C (-4 to 158 °F) 95% relative humidity, non-condensing
1 second to 1 minute, adjustable average
> 1 L/min, 2-4 L/min recommended
7 to 35 °C (45 to 95 °F)
½-in. BSP
RS485 electrically isolated; data rate: 1200 to 9600 baud
4-20 mA card: Two 4-20 mA outputs; maximum load 500 Ω, electrically isolated
Alarm relay card: Two relays; SPST, NO and NC terminals, maximum 16 VAC
(35 VDC), 5 A (resistive loads only)
Two dual digital; optically isolated; switch input or 5 to 24 V input
CE, cETLus (UL and CSA 61010-1 safety standards)
1 year